The MIKE (Magellan Inamori Kyocera Echelle) User's Guide
Rebecca Bernstein, Steve Shectman, Steve Gunnels, Ian Thompson, Greg
Burley,
Christoph Birk, Stefan Mochnacki, Alex Athey
This manual contains the basic information needed to observe with
MIKE. The instrument is described in more detail in Bernstein, Shectman, Gunnels, Mochnacki,
Athey 2002, Proc SPIE 4841. For occasional updates
and discussions regarding MIKE, the Magellan telescopes, and other
Magellan instruments, a "News &
Forums" page is maintained at LCO.
1. Overview:
MIKE is a double echelle
spectrograph. The first optical element in the spectrograph
is a
dichroic which reflects (transmits) light into the blue (red) arms of
the spectrograph. Each side has its own shutter and CCD, so the
two sides can be used
simultaneously with independent exposure times. The basic
specifications of each side are listed in the table below.
The parameters which the observer can select are exact wavelength
coverage and the slit width.
The spectrograph delivers full wavelength coverage from about 3350-5000
(blue) and 4900-9500A (red) in its standard configuration.
This configuration is set by the cross over wavelength of the dichroic
(4950A).
As the observer, you can choose to have the grating angle adjusted to
select a different set of orders to go bluer on the blue side (down to
3200A) or redder on the red side (up to 10,000A). But it is not
possible to use a different dichroic, gratings, or prisms. These
elements are permanent.
Because MIKE uses prisms for cross dispersion, the separation between
orders changes as a function of wavelength. The smallest
separation is about 6". The longest slit that can be used for
full wavelength coverage is therefore 5" long. Several different
slit widths are available, as discussed below. Choose a slit
width that gives you the resolution you want.
A small blue box in the control room allows the observer to remotely
control internal lamps (ThAr and incandescent) , the slit
position, and the focus of the cameras. The first time user
should let the Instrument Specialist focus the spectrograph.
The CCDs and their electronics were provided and are maintained by Ian
Thompson and Greg Burley. All information related to upgrades,
linearity, and detailed behavior has been provided by Ian Thompson.
MIKE also has a fiber system which was built by Mario Mateo (UM) and
Alex Athey (UM, OCIW). The fiber-fed mode is not yet
available for use by guest observers without explicit permission and
assistance from Mateo.
2. Updates:
October 2005:
1- Ian Thompson reports that he has fixed the linearity problems
associated with the blue side CCD (Lincoln Labs, installed May
2004). Any data taken after September 20, 2005 should
be linear to digital saturation of the A/D
converter.
May 2004:
1- A new dichroic was installed which has a crossover
wavelength of ~495 nm. (The old dichroic produced a crossover
wavelength of ~455nm.)
2- A Lincoln Labs CCD was
installed on the blue side. Like the original, this CCD has 2k x 4k x
15 micron pixels. On the plus side, it has very high quantum
efficiency. On the minus side, it
has limited
dynamic range. This is discussed further in Section 4
("Efficiency and Exposure Times"), below. The bottom line is that
you should keep the total count levels in your data below 8,000 DN per
physical pixel (1x1 binning), and below 16,000 DN at any
binning. [As of October 2005, this problem has been fixed,
as noted above.]
3. Basics:
|
Blue
Side
|
Red
Side
|
effective
focal ratio
|
f/3.9
|
f/3.6
|
scale
at CCD
|
8.2
pix/" (0.12"/pix) |
7.5
pix/sec (0.13"/pix) |
Å/pixel
(unbinned)
|
~0.02
|
~0.05 |
|
|
|
detector
|
2048x4096
(15µm pix) |
2048x4096
(15µm pix) |
| gain |
~0.47 e-/DN
|
~1.0 e-/DN
|
Readnoise
|
~2 e-/pix
|
~3.5e-/pix
|
Dark current
|
~5 DN/pix/hr
|
~2 DN/pix/hr
|
CCD efficiency
|
|
Q.E.
|
|
|
|
Wavelength
range*
|
3200
– 5000 Å |
4900
– 10000Å |
Resolution
(0.35"slit)
|
83,000
|
65,000
|
Resolution
(1.0"
slit)
|
28,000
|
22,000
|
|
|
|
Echelle
grating
|
R2.4
|
R2
|
Prism
(cross-disperser)
|
Fused
Silica (2 prisms) |
PBM2
(1 prism)
|
* The grating position and wavelength
ranges for the standard setup are given in Section 8.
4. Resolution and binning:
The cameras make very
good images and can easily
resolve a 0.35" slit (the smallest slit available). That means
that the number of pixels per resolution element of your spectra will
be limited only by the slit width
you use (smaller slits mean higher resolution, linearly related).
If you are observing very faint objects,
it pays to bin the data in the spectral and spatial
directions.
The scale of the CCD detectors is about 8.2 pixels/" (blue side) and
7.5 pixels/" (red side). So if you are
using a slit which is wider than 0.5 arcseconds there is not much point
in taking unbinned data. In other words, you should bin by
in the spectral (y) direction so that slit width is Nyquist
sampled. In addition to significant gains in readout time, your
data will be less effected by readnoise, which does not increase
significantly with binning. It is very
common to use MIKE binned 2x2 or 3x2 on both sides.
5. Efficiency and Exposure Times:
The plot below shows the source flux (in AB mag)
from
which we detect 1 e-/sec/A. The slit-to-detector efficiency is
roughly
37% (~4500) on the blue side and 20% (~6500A)
on the red side.
(Measurement
and calibration courtesy of Scott Burles and Kristin Burgess, MIT.)
This plot is based on data taken through a 2" slit
in 0.7" seeing, so not much light was lost at the slit. If
you use a slit width which is matched to the seeing (1" slit in 1"
seeing), you should expect to lose about 25% of the light at the
slit.
This plot is all you need to estimate the exposure
times for your program. To be clear, here is an example of how to
use
this plot:
Flux of target star:
V=15 mag
Desired SNR:
50 per pixel (i.e. ~ 2500
e-/pixel)
# Angstroms/pixel:
0.05 A/pixel red side, w/o
binning in spectral
direction (y)
seeing, slit, airmass:
seeing=0.8" ,
slit=0.7" (% into slit relative to plot ~ 0.7), airmass=1.0
count rate ~5500A:
1 e-/sec/A * 0.05 A/pixel *
10^(0.4*(18.4-15)) * 0.7 = 0.77 e-/pixel/sec
required exp. time: 2500 e-/pixel /
0.77 e-/pixel/sec = 3250 sec.
Readnoise per pixel is small (4-5 DN/pixel), so this
simple estimate is
approximately correct. Remember that this calculation has to do
with
the integrated flux over the spatial extent of the source (i.e. peak
flux in the spatial direction near 1000 could correspond to a total
flux of 2500 DN integrated over all pixels at the same wavelength once
you
extract the spectrum). Also remember
that this calculation assumed 0.8" seeing and an airmass is near
1. You should adjust
accordingly if you
have greater slit loses (bad seeing) or are at high airmass.
If you are observing faint objects that are going to give you a
relatively small number of counts per pixel in your chosen exposure
times, then readnoise will obviously be more important. In this
case,
you should definitely opt for on-chip binning. As
we discuss below, the scale of the CCD detectors is 8.2 pix/"
(blue) and 7.5 pix/" (red). So if you are
using a slit which is wider than 0.5", you should bin to minimize
amplifier noise. For example:
Flux of target star:
V=18 mag
Desired
SNR: 20
per pixel
(i.e. ~ 400 e-/pixel)
# Angstroms/pixel:
0.1 A/pixel red side, w/ x2
binning in spectral
direction (y)
seeing, slit, airmass:
seeing=0.6", slit=0.7"
(% into slit relative to plot ~ 95%), airmass=1
count rate ~5500A:
1 e-/sec/A * 0.1 A/pixel *
10^(0.4*(18.4-18)) * 0.95 = 0.14 e-/pixel/sec
required exp. time: 400 e-/pixel /
0.14 e-/pixel/sec = 46 min.
CCD Linearity:
Data taken with both the red and blue sides
should be linear up to the digital saturation of the A/D converters,
with the exception of blue-side data taken between May 2004 and Sept
20, 2005 (using Lincoln Labs CCD, installed May 2004).
Until Sept 20, 2005, two linearity
problems affected the Lincoln Labs CCD that was installed in May
2004. These problems were measured and reported by Ian
Thompson. First, for reasons apparently having to do with the
electronics, individual pixels appeared to saturate at about 8,000
DN. For all unbinned observations (1x1
readout), saturation set in around 8,000 DN. A
discussion of this problem (the full-well limit) was posted by
Dave Osip to the "News & Forums" page (copied here). Second, the amplifiers
appear to saturate at about 16,000 DN.
For
all binned observations, saturation appeared to set in at around 16,000 DN. A plot
showing the effect of the amplifier saturation is available here (plot, Ian
Thompson). As of Sept 2005, Ian reports that both of
these problems have been fixed.
6. Lamps:
A. Internal comparison source:
MIKE has an internal Thorium Argon lamp. The vast majority of the
time, and certainly for the standard set-up, this is what you want to
use for arcs. It can be turned on and off from the remote control
box in the data
room.
The switch labeled "comparison lamp" will both turn on the lamp and
move
a flipper-mounted mirror into place to direct the light onto the
slit.
A one second exposure is okay for the blue side and most of the red
side. You will notice that there are strong
lines which become very saturated in the orders around 7500A. If
you display the images with the right stretch, you'll see that the
saturation isn't as bad as it looks. These regions do not pose a
calibration problem. However, you may also notice that the
reddest
10 orders or so have fairly few strong lines. This is a bigger
problem. Be sure to look closely at these orders to be sure
you're getting the counts you need. When binned 2x2, a 1 sec
exposure is adequate, but a 5 sec exposure gives you more lines.
Wavelength and order maps can be found on the main MIKE web site
at LCO.
It is very important not to leave the internal ThAr lamp on when it is
not needed because it has a lifetime of only a few hundred hours
and external sources will not be bright enough for blue-side
calibration.
The required exposure time is nice and short, so please get into the
habit
of turning it off promptly when the exposure shutters close. It
is not possible to
accidentally leave the arc lamp on during an exposure because the arc
light floods the slit viewer with light. It's hard to ignore and
it is impossible to place an object on the slit in this state.
B. External comparison source:
It is also possible to use external
calibration lamps for the red side, although it is extremely unusual to
use these and
probably not
necessary. If you want to use them for some reason (for above
9500A?), you need to move the flat
field screen into
place using the GUI shown below. To start the GUI, type "ffs" in a
xterm
on the data computer.

The row of green boxes represent individual arc lamps mounted near
the
screen. The available lamps are Neon, Argon, and Helium. Ar
+ Ne together work well for the red orders. Click on one of the
green
buttons to turn on the lamp. The button will turn red when the
lamp
is on. Click it again to turn it off. Below the row of arc
lamps is a row of boxes that move the flat field screen in and out of
the
telescope field of view. When the screen is in, the button in the
lower right corner will be red and read "In." A few
seconds
is adequate for the red side. Use the internal comparison source
for the blue side.
C. Internal incandescent Lamp:
There is also an internal flat field lamp which is controlled from the
box
in the control room. You can turn on this lamp simply by flipping
the arc-lamp switch down instead of up. The label of the
switch indicates this. This lamp works very well for flats on the
red side, with or without the diffuser slide (milky flat slide).
It is not very bright on the blue side. Methods for taking flats
are discussed further below (Section 7c).
7. Taking data:
A. Selecting a slit:
MIKE has a polished reflecting slit plate with a variety of slits
machined
into it. Behind the slit plate is a fixed plate which blocks the
light from all of the slits except the one which is positioned in front
of a single hole at the center of the field. The slit plate is mounted
on a motorized stage so that any of the slits can be positioned in
front
of the single hole in the blocking plate. The telescope operator
will know where the blocking plate is located on the slit viewing
camera. He will mark on the slit viewer the position where the
desired slit should be located for observing and will help you position
the slit. The pixel scale on the slit viewing camera is
0.067" per pixel, so you can see the position of the slit with good
resolution.
The easiest way to select a slit during the day is by using the
internal flat field lamp to illuminate the slit plate. Because
this is
an internal lamp, the state of the telescope is irrelevant (i.e. the
mirror covers can be closed). The
lamp will flood the slit viewing camera with very bright light, and so
it will need to be adjusted for large dynamic range (using the "span"
command on the slit viewing camera). It is basically not possible
to do this without a little experience, so you should ask the
Instrument Specialist or Telescope Operator for
help the selecting the slit you want to use during the afternoon at the
start of
your run. Once you can see the slits in the slit viewer, just use
the switches on the control box
to move the slit plate to the right or left and select a
slit. The switches are well labeled. There
is a handy knob for controlling the speed of the slit motion.
There is no encoder to identify the exact position of the slit
plate. We are working on this, however it is relatively easy to
move the slit and return it to within 1 pixel in the dispersion
direction just by eye in the slit viewing camera (see
below). Because
standard arcs tell you the wavelength position of the spectra, it is
not a problem to move the slit between exposures or use multiple slits
during a run. Arcs should be taken every 30-60 minutes, anyway,
as discussed below.
This is what you will see in the slit viewer (left) and a schematic
drawing of the slit viewer plate (right):
Available slits from left to right on the slit viewer screen are as
follows:
Aperture Pairs (separation 3"):
1 0.35 x 0.35 (for focusing)
2 1.00 x 0.35
3 1.00 x 0.50
4 1.00 x 0.70
5 1.00 x 1.00
6 1.50 x 0.35
7 1.50 x 0.50
8 1.50 x 0.70
9 1.50 x 1.00
10 1.50 x 1.50
11 2.00 x 0.35
12 2.00 x 0.50
13 2.00 x 0.70
14 2.00 x 1.00
15 2.00 x 1.50
16 2.00 x 2.00
Single Slits:
17 0.35 x 5.00
18 0.50 x 5.00
19 0.70 x 5.00
20 1.00 x 5.00
21 1.50 x 5.00
22 2.00 x 5.00
Most of the available slits are in fact pairs of apertures. You
can use these to observe in "A-B" mode, in which you would obtain one
exposure with the star in the top
aperture followed by a second exposure with the star in the bottom
aperture. One aperture is then "object" and one is "sky" in
the "A" exposure, and swapped in the "B" exposure. An
easily-identified spatial portion of the slit is dedicated to sky and
the object this way. This lets you bin the data very
aggressively in the cross-dispersion (x) direction; you can bin
by 4 with the 2 arcsecond apertures. Binning is limited by the 1"
gap between the slits. Because the 1.5"-long aperture pairs are
1.5" apart, you can bin be larger values with these apertures.
A-B mode is especially useful when
observing very faint objects because it is possible to locate and sky
subtract very
faint objects this way. On the other hand, MIKE does not
currently have an atmospheric dispersion corrector. (We are in
the
process of adding one.) So if you are observing at airmasses
higher than 1.2, you will have some light at
blue wavelengths dispersed out of the short apertures. Also, if
you are working an a crowded field, it may not be possible to use the
paired apertures and always get a clean measurement of sky.
B. Wavelength Calibration:
The internal ThAr lamp described above (see "Lamps") can be used
to take all the arcs you should need during your run.
We often bin by 2 in y (spectral direction) on both sides, and
find that a 1 sec exposure gives adequate flux (even at >7600A) to
obtain wavelength solutions with residuals <0.04 pix rms with a
standard IRAF reduction. However, we find 5 sec arcs give us more
lines to work with at the red end. We therefore recommend about a
5-10 sec exposure time for the internal ThAr arcs depending on your
binning.
An arc line map is available at on the MIKE web site
at LCO.
C. Flat fields:
Because this is an echelle, different colors of light are imaged on the
slit in different areas. You want a flat that changes color with
position. You wouldn't want to take a flat without the slit in
place. But you do want to fill the CCD with light (even
between orders!) in order to get a good flat field image for
pixel-to-pixel sensitivity corrections. To do this, a
diffusing glass slide can be positioned in the optical path just
downstream
of the slit for taking "milky flats." This "milky flat" slide
will blur the slit image into a roughly 40-50 pixel smudge. A
40-50 pixel smudge is a good size because it means
that there is not much light shared between orders in the flat, but the
gaps are well illuminated. This lets you flat field at the edges of the
orders.
The diffusing slide is mounted on a sliding post and can be positioned
in or out of the optical beam manually. The post
sticks through the blue adapter plate below and left of the blue dewar
(between the blue and red dewars) and has a brass knob on the
end. When the knob is pushed in (flush with the adapter
plate), the
diffuser
is
out of the beam. When the slide is pulled out,
the diffuser is in position for milky flats. There is a detent at each
position. Push and pull firmly and slowly. The brass
knob is well labeled.
To take good flats, you need a light source which supplies a lot
of photons and does not have strong spectral features. The
internal incandescent lamp is fine on the red side of MIKE, but it does
not supply much light for the blue side. For that reason,
we STRONGLY suggest that you take some milky flats during twilight
using an O or B star. We have found that a star with V=4-5 mag
gives very good flats in about 25 seconds per exposure. It is a good
idea
to take 1/2 of the flats with the star at one end of the slit, and 1/2
with the star at the other end of the slit. That way, the
inter-order region is better illuminated than if you only position the
star in the center of the slit. The best stars for this
purpose are stars with high rotational velocities -- high
v*sin(i).
We have left a catalog of such stars with the Telescope
Operators ("high_vsini_obscat.cat"). The stars on this list
have high enough Doppler
shifts that all spectral features are very broad and can easily be
median smoothed out with a 1x31 kernel to produce a very nice flat
field image. It is relatively easy to take 10 exposures in
twilight per night. They
can be combined over a run if you have multiple nights. We
have found the flats to be pretty stable.
D. Atmospheric
Dispersion
MIKE was designed to be mounted at the Nasmyth or auxiliary
port. We expect that the preferred observing strategy will
be to use MIKE in a gravity invariant mode in which it is mounted on
the East Nasmyth platform and does not rotate with the instrument
rotator. This is the only mode
available now. The virtue of this mode is obviously the
stability of the instrument. The down side is that the slit
orientation on the sky cannot be changed. The only potential difficulty
this poses for point sources has to do with atmospheric dispersion away
from zenith. For that reason, we
have positioned MIKE at a 30 angle to the platform. At this
angle,
atmospheric dispersion lies exactly along the slit when observing at a
zenith angle of 30 deg.
The figure below shows the atmospheric dispersion (3500-9000A) across
and along the slit as a function of zenith angle for MIKE in its
standard position (tilted by 30 degrees) on the Nasmyth
platform. At a zenith angle of 30 degrees, the
dispersion is entirely along the slit with a magnitude of about 1.1
arcsec. The dispersion across the slit is small
(<0.3 arcsec) for all zenith angles less than 40 deg. At
higher zenith angles (>50 deg), atmospheric dispersion across the
slit is significant. With the 5"–long slits, most of
the light should get into the slit at zenith angles less than 60
deg. If you are using one of the aperture pairs (2"–long slits or
shorter) and are
interested in the full wavelength range, then it is probably better to
avoid zenith angles greater than about 40 deg.

E. Recommendations regarding calibration:
Bias: Occasional jumps in the bias level occur but can be
subtracted
out nicely by using an average of the overscan region at the end of
each
row (x=2048:2176,y=*, unbinned). The highly repeatable bias ramp
which occurs at the beginning of each row can be easily subtracted
using
the overscan at the end of each column (x=*, y=4096:4224,
unbinned).
Using these overscan regions is more effective than an using a combined
bias frame. For that reason, 0-second exposures (bias frames) are
not very useful and are not recommended.
Dark: The new Lincoln Labs blue CCD (installed May 2004)
has some dark current (~5 DN/hr, depending on running
temperature)
which will add noise to the exposure but does not appear to have
significant
2-d structure over the CCD. Check the dark level during
your
run and keep this in mind when estimating the signal to
noise
ratio of your final exposure.
Arcs: It is a good idea to take arcs between long
exposures
(every 15-60 minutes). Although there is no motion in the 2-D
spectrum due
to flexure (the instrument doesn't move), there is motion due to
temperature changes of the air,
glass, and metal. Over the course of a night, temperature changes
can cause small (~ 1pixel) shifts, mostly in the cross-dispersion
direction. So the telescope pointing doesn't
affect the wavelength calibration, but time does! It is fine to
move
the telescope while the data is reading out. It is also fine to
take an arc while moving between objects.
Flats: It is a very good idea to take flats with the diffuser
for standard
pixel-to-pixel flat fielding calibration. You also want to
take
some flats (either internal lamp or sky flats 0-15 min before sunset)
without the diffuser. These can be used for an illumination correction
and to trace the edges of the orders. Take an arc at the same
time
to provide an easy way of identifying the offset between the order
position
in the flat and the order position in each program
exposure.
(The orders can move by fractions of a pixel in both the dispersion
direction
and the cross-dispersion direction due to temperature changes in the
spectrograph.)
8. Summary of moving parts:
A. Electronically controlled
A control box is mounted on the side of MIKE for "local"
control.
A second, remote box is located in the telescope control
room.
All of the controls are manual switches or knobs and are
labeled.
Be sure that the local control box switches are set to "REMOTE" so that
the slit, comparison, and focus can be controlled from the data room.
1. Slit plate:
The position of the slit is controlled by a motorized stage and
positioned
using the internal slit-viewing camera as described above (see Taking
Data). You can move the slit plate at will to select the slit you
want during the night.
2. Internal lamps (Th+Ar lamp and Incandescent lamp)
The lamp switch also controls the flipper mirror as described
above (see Comparison Lamps) . You will need to take arcs
throughout your observation at night and use the Incandescent lamp for
flats during the day.
3. Camera Focus:
The Instrument Specialist will focus MIKE before your run. You
should not need to adjust the focus of the spectrograph during your run.
So this is for your
information
only.
Each camera is mounted on an optical bench inside MIKE. The
position of each camera relative to the CCD is controlled by a
motor-driven cam which moves the whole bench relative to the
CCD. These drives are controlled remotely from the box in
the control room to focus the cameras. The blue camera focuses at a
setting of 1.150. The red camera focuses
around 1.050. The optical benches also each contain an invar plate
which will move the camera as the aluminum expands or contracts with
temperature. This passive thermal control keeps MIKE in good
focus over temperature changes as large as 5-10 degrees without any
adjustments to the motorized cams. Empirically, the focus is
extremely
stable. In fact, the best-focus settings do not change
appreciably even on 1 year time scales (winter to summer). If you
try to refocus the instrument on any given day, you may find a slight
change from the previous day if the temperature has changed by more
than 5 degrees. However, you are probably just chasing temporary
thermal mis-matches in the metal and glass from day to night.
These are probably changing fairly consistently between when you focus
(afternoon) and when you observe (night). It is not recommended
that you try to refocus MIKE for temperature changes smaller than 10
degrees or if the temperature fluctuating through normal 12 hour time
scales. If you do try to refocus, try moving in steps of
0.010.
B. Manually controlled:
1. Diffusing slide for "milky flats". You will need to
move this slide into and out of the beam as necessary during your run
to take milky flats.
2. Grating position
The Instrument Specialist will set the grating positions. We
do not recommend doing this yourself. If you want a different
wavelength range, please specify this in your instrument set-up
form. This section is for your information only.
The gratings are positioned manually against preloaded, threaded
rods.
Dial gauges underneath the red and blue boxes indicate the linear
position
of the rods, and therefore the position of the grating. There should be
no need to move the blue grating. The azimuthal position of
the red grating can be adjusted to select the orders which fall on the
chip. Each of the gauges and threaded rods are labeled on the
spectrograph.
As of May 2004, the standard grating settings and corresponding
wavelength
coverage are as follows:
Blue Azimuth: 0.346 (was 0.374)
Red
Azimuth: 0.474 (was
0.535)
Blue Elevation: 0.540 (same)
Red Elevation: 0.380 (same)
Orders 71-106
Orders
37 - 70
5000 - 3350 A
9300-4900 A
These elevation settings center the free spectral range well, and
should not be changed. The azimuthal settings can be adjusted to
change the order coverage.
The images below are of twilight sky about 10 minutes before
sunset. Blue is on the left. Both images have (x,y)=(0,0)
at the lower left. Because MIKE uses prisms for cross dispersion,
the redder orders are
always closer together than the blue. This makes it very easy to
get your bearings in general. The MIKE web page at LCO also
includes arc maps and images with orders labeled. The
strong solar features are easy to identify, as are the atmospheric
bands. (CaII H&K are the huge absorption features in the same
order near the center of the blue image. The atmospheric A and B
bands are the strongest features on the red side.) On the blue
side, the spectra get redder towards x=small (on the left) and y=large
(at the top). On the red, spectra get redder towards
x=large (on the right)
and y= small (on the
bottom).


9. MIKE data acquisition GUI
The data control window is shown below. To start the window, type
"mike"
in an xterm on the data control computer. An instrument configuration
window
will appear, also shown below. The observer can enter his/her name and
then select three camera configurations, blue camera only, read camera
only, or both blue and red camera. There is an entry for off-line
(simulator)
operation of the GUI, this is for testing purposes only. Next the
observer
should select the number of pixels of overscan at the end of each row
and
the number of bias lines at the end of a readout that should be
appended
to a frame. (128 in x and y is good.) Finally the observer
can select the telescope, and again an
online or offline operation. If the telescope is online then telescope
coordinates and various rotator angles are read from the telescope TCS.


The very top portion of the acquisition window has two pull down
menus,
the first controls the window status (select Exit GUI to exit the
window)
the second has an entry for user name and another to select the path to
the data. This path can be changed without having to restart the
window. (You can set up any directory structure you like under
~/DATA/.) These are followed by a bar graph showing the current
data disk
capacity,
and a display of the current UT.
The top half of the window shows information relevant to BOTH the
red
and blue side, like telescope coordinates, dome temperatures, airmass,
etc. Exposures for simultaneous operation of both cameras with all
exposure
parameters identical are controlled by the Exptime, Start, Pause,
Abort,
and Loop buttons in the upper right portion of the top half of the GUI.
The temperatures of the CCD's are given in the top half of the window.
The current set points are -125 C, and these numbers will turn red if
the
temperature rises above the set point by more than 5 C.
Entries of numerical values for image number and exposure time will
cause the background in these windows to turn red, the values are
entered
into the program when you hit the return key. Other character fields
(white
boxes for comments, settings for camera focus values, grating angles,
slit-size)
can be simply edited. Entries for X binning, Y binning, full or
subrastered
readout, and speed are controlled with pop up menus, indicated by the
presence
of a small circle in the right side of a menu. Two readout speeds are
available,
a fast, high-noise mode (full frame readout 95 seconds, readnoise
approximately
4.7 electrons), and a slow, low-noise mode (full frame readout 160
seconds,
readnoise approximately 3.7 electrons). Note that the read times in all
cases will be significantly shorter when a camera readout is binned.
When
exposures are running, the start buttons turn yellow in the exposing
cameras.
Start/Pause/Abort buttons are beige when those functions are available.
Cameras can also be operated together using the top half of the
window or completely independently with the control
buttons in the lower left (blue) and lower right (red) portion of the
window.
Exposure times can be different, exposures can be started
asynchronously,
and the number of exposures in a loop can be different.
Note also that exposure time can be
changed during an exposure! Just move the cursor over the
"Exp. Time" box in the top half of the window, type a new number, and
hit return. This is the easiest way to abort an
exposures. You can abort the readout also, if you like, using the
abort button.
There are two message windows at the bottom of the GUI. These give
the
status of an individual readout, both the elapsed read time and the
number
of lines read and the total number of lines to be read.
The data taking system is pretty robust at this point. We don't
know of anything you should particularly avoid doing. If you do
find a way to break the GUI, please tell the Telescope Operator and
make sure to include it in the Observer's Report.
10. Random (very incomplete) Notes for Instrument Specialists:
1- Make sure that of the necessary changes have been made to the guider
and Shack-Hartmann parameters before and after the fibers are used with
MIKE. This is normally done by the Telescope Operators, but it
can be forgotten between crew shifts.
2- The control box on the side of
the instrument has a knob which controls
the brightness of the comparison lamp. Turning the knob clockwise
makes the lamp brighter. At any setting, the lamp is already too
bright, but the power supply doesn't regulate properly at the lowest
setting.
There is a mark on the box at or above which the power supply works
properly.
Leave the knob set to the mark.
3- If the control box in the observing room doesn't seem to be working,
make sure that the control switch in the box mounted on the side of
MIKE is set to REMOTE.
Last updated by rabernst at UM, May
2005.