Problem: Agilent Not Reading Temperatures
Manifestation:
-
Agilent not reading any temperatures, hence
ShARCS detector temperatures cannot be monitored.
Power-cycling device unsuccessful.
Display reads MUX OPEN 102.
Solution/Recovery:
-
View Agilent configuration via web browser at the following URL:
ao-agilent.ucolick.org
-
Select the second icon down on the left to start the IcedTea
java application to control the Agilent.
-
Select: Allow Control and set up monitoring for channels
102, 103, 105, 107, 108, and 121 by right clicking on each
of those channels.
-
Re-start sharcstherm service:
ao stop sharcstherm
ao status sharcstherm
ao start sharcstherm
Cause:
-
Unstable behaviour following power-glitch/power-cycle.
Associated Issues:
Change Log:
- 2017-02-10: First version.
|