Problem: laserOffload GUI Freezes
Manifestation:
laserOffload GUI unresponsive. Values not updating.
This occurs regardless the activity of TT loop, Uplink TT loop and WFS loop.
Solution/Recovery:
1st Attempt
-
In ShaneAO loop control GUI, cycle Uplink: On-Off-On.
2nd Attempt
-
In slinelamp_fe GUI, cycle Spare2: On-Off-On.
3rd Attempt
Restart shaneao software.
-
In ShaneAO terminal window, hit enter key to acquire python prompt >>>
-
At prompt, issue the following command: exit()
-
ShaneAO loop control GUI should modify to a yellow dispatcher
warning.
-
At prompt, issue the following command: shaneao
-
ShaneAO loop control GUI should return.
Cause:
Unknown.
Associated Issues:
None
Change Log:
- 2016-11-16: Updated (Cycle Spare2 power supply).
- 2015-11-29: First version.
|