Shane AO  

Welcome/Index | Architecture | Procedures | Problems | Features | Documents | Resources | Chronology  



SHARCS Problem: Ion Pump Shut Down

Manifestation:

  • ShARCS ion pump control () reports HV power Off.
  • ShARCS ion pump trips Off after short period of operation.
  • In response to following command:

    user@covert > gshow -s sharcspump -date "48 hours ago" PRESSURE HVPOWER

    HVPOWER = Off

  • Ion pump controller (mounted external to ShaneAO) displays message:Sd0 and all other indicators are off.

Solution/Recovery (Temporary):

  • Power-cycle ion pump controller (multiple times, if necessary):
    • Visit ion pump controller (mounted external to ShaneAO).
    • For completeness, press HV button to disable High voltage supply.
    • Throw rocker switch on unit to power OFF position for 30 seconds.
    • Throw rocker switch on unit to power ON position.
    • Wait a few seconds for power-up sequence to complete.
    • When stable after power-up, press and hold HV button for 2 seconds.
    • Stand-by for circa 5 minutes to observe beaviour and verify stability.

Solution/Recovery:

  • HV cable: Inspect, examine, repair/replace.

Cause:

  • Faulty High Voltage cable / BNC connector.

Associated Issues:

Change Log:

2017-05-10: First version.


This document last updated (UTC): Tuesday 04 March 2025