Procedure: ShARCS Calibration
Spectroscopic Wavelength Calibration
Spectroscopic Flat Field Calibration
Imaging Flat Field Calibration
Spectroscopic Wavelength Calibration
-
Install Argon arc lamp (manually).
-
Set Cal X and Cal Y to White Light
-
Use Cal X and Cal Y detail panels realtive move
with values of a few 100 encoder counts to maximize WFS illumination
-
Configure filter and aperture wheels for K-band spectroscopy:
Filter#1 = CaF-Kgrism and Filter#2 = K and Aperture = Slit-100mm-H
For Slit-100um-H, the target encoder value is: 0 ± 4.
-
Set Window to Full Frame.
Set NCoadds to 1.
Expose a minimum of 3 arc lamp spectra,
seeking to achieve a few hundred to a few thousand counts
(not exceeding 20000 counts) in several arc lines.
Example exposure times (2015) in range 1.5 to 15.0 seconds.
Record file, configuration, exposure time.
-
Configure filter and aperture wheels for H-band spectroscopy:
Filter#1 = H and Filter#2 = HGrism and Aperture = Slit-100mm-H
For Slit-100um-H, the target encoder value is: 0 ± 4.
-
Set Window to Full Frame.
Set NCoadds to 1.
Expose a minimum of 3 arc lamp spectra,
seeking to achieve a few hundred to a few thousand counts
(not exceeding 20000 counts) in several lines.
Example exposure times (2015) in range 1.5 to 6.0 seconds.
Record file, configuration, exposure time.
-
Set Cal Y to Open
-
Remove Argon lamp (manually).
Spectroscopic Flat Field Calibration
-
Set Cal X to White Light
-
Set Cal Y to Open
-
Set sdiagmir_fe to position 3
-
Ensure ambient enclosure illumination is OFF
and primary mirror cover OPEN
-
In slinelamp_fe set Sup_Blue Dimmer to 25
-
Configure filter and aperture wheels for H-band spectroscopy:
Filter#1 = H and Filter#2 = HGrism and Aperture = Slit-100mm-H
For Slit-100um-H, the target encoder value is: 0 ± 4.
-
Set Window to Full Frame.
Set NCoadds to 1.
Expose a minimum of 3 spectroscopic flat field frame,
seeking to achieve a few hundred to a few thousand counts
(not exceeding 20000 counts) in several lines.
Example exposure times (2015) circa 18.0 seconds.
Record file, configuration, exposure time.
-
Configure filter and aperture wheels for K-band spectroscopy:
Filter#1 = CaF-Kgrism and Filter#2 = K and Aperture = Slit-100mm-H
For Slit-100um-H, the target encoder value is: 0 ± 4.
-
Set Window to Full Frame.
Set NCoadds to 1.
Expose a minimum of 3 spectroscopic flat field frame,
seeking to achieve a few hundred to a few thousand counts
(not exceeding 20000 counts) in several lines.
Example exposure times (2015) circa 18.0 seconds.
Record file, configuration, exposure time.
-
In slinelamp_fe set Sup_Blue Dimmer to OFF
Imaging Flat Field Calibration
-
Set Cal X to White Light
-
Set Cal Y to Open
-
Set sdiagmir_fe to position 3
-
Ensure ambient enclosure illumination is OFF and primary mirror cover OPEN
-
In slinelamp_fe set Red to On
-
Configure filter and aperture wheels for (typically) Ks-band imaging:
Filter#1 = Ks and Filter#2 = Open and Aperture = Open
-
Set Window to Full Frame.
Set NCoadds to 1.
Expose a minimum of 3 flat fields,
seeking to achieve a few thousand counts
(not exceeding circa 20000 counts).
Example exposure times (2015) circa 12.0 seconds.
Record file, configuration, exposure time.
-
Configure filter and aperture wheels for (typically) H-band imaging:
Filter#1 = H and Filter#2 = Open and Aperture = Open
-
Set Window to Full Frame.
Set NCoadds to 1.
Expose a minimum of 3 flat fields,
seeking to achieve a few thousand counts
(not exceeding circa 20000 counts).
Example exposure times (2015) circa 12.0 seconds.
Record file, configuration, exposure time.
-
Configure filter and aperture wheels for (typically) J-band imaging:
Filter#1 = J and Filter#2 = Open and Aperture = Open
-
Set Window to Full Frame.
Set NCoadds to 1.
Expose a minimum of 3 flat fields,
seeking to achieve a few thousand counts
(not exceeding circa 20000 counts).
Example exposure times (2015) circa 40.0 seconds.
Record file, configuration, exposure time.
-
In slinelamp_fe set Red to Off
|